# boulderopalscaleup.experiments.ResonatorSpectroscopyByPower

Source: https://docs.q-ctrl.com/references/boulder-opal-scale-up/boulderopalscaleup/experiments/ResonatorSpectroscopyByPower

```python
class boulderopalscaleup.experiments.ResonatorSpectroscopyByPower(*, resonator, frequencies=None, powers_dbm=None, duration_ns=2000, recycle_delay_ns=1000, shot_count=100, run_mixer_calibration=False, update='auto')
```

Parameters for running a resonator spectroscopy by power experiment.

### Variables
* **resonator** (*str*) – The reference for the resonator to target.
* **frequencies** ([*LinspaceIterable*](https://docs.q-ctrl.com/references/boulder-opal-scale-up/boulderopalscaleup/common/LinspaceIterable.md#boulderopalscaleup.common.LinspaceIterable) *or* [*RangeIterable*](https://docs.q-ctrl.com/references/boulder-opal-scale-up/boulderopalscaleup/common/RangeIterable.md#boulderopalscaleup.common.RangeIterable) *or* [*CWSIterable*](https://docs.q-ctrl.com/references/boulder-opal-scale-up/boulderopalscaleup/common/CWSIterable.md#boulderopalscaleup.common.CWSIterable) *or* *None* *,* *optional*) – The frequencies at which to scan, in Hz.
  Defaults to a scan around the readout frequency.
* **powers_dbm** ([*LinspaceIterable*](https://docs.q-ctrl.com/references/boulder-opal-scale-up/boulderopalscaleup/common/LinspaceIterable.md#boulderopalscaleup.common.LinspaceIterable) *or* [*RangeIterable*](https://docs.q-ctrl.com/references/boulder-opal-scale-up/boulderopalscaleup/common/RangeIterable.md#boulderopalscaleup.common.RangeIterable) *or* *None* *,* *optional*) – The powers at which to scan the readout pulse, in dBm.
  If None, a default power scan will be used.
* **duration_ns** (*int* *,* *optional*) – The duration of the readout pulse, in nanoseconds. Defaults to 2,000 ns.
* **recycle_delay_ns** (*int* *,* *optional*) – The delay between consecutive shots, in nanoseconds. Defaults to 1,000 ns.
* **shot_count** (*int* *,* *optional*) – The number of shots to take. Defaults to 100.
* **run_mixer_calibration** (*bool* *,* *optional*) – Whether to run mixer calibrations before running a program. Defaults to False.
* **update** ( *"auto"* *or*  *"off"* *or*  *"prompt"* *,* *optional*) – How the device should be updated after an experiment run. Defaults to auto.